Keysight (Agilent) 84906K

Programmable Step Attenuator, DC to 26.5 GHz, 90 dB, 10 dB steps
Contact us for pricing and availability.
Description
The Agilent 84906K programmable step attenuator offers coaxial measurements to 26.5 GHz in a compact, rugged design. It offers repeatability of better than 0.03 dB, excellent life (greater than 5 million switching cycles per section) and an attenuation range of 0 to 90 dB in 10-dB steps. This latest design evolution sets new standards for size and performance. High attenuation accuracy and low SWR are achieved through the use of miniature thin-film attenuation cards composed of high-stability tantalum nitride film of saphire substrate. Insertion loss is outstanding, with less than 2 dB of loss at 26.5 GHz. The compact size of the unit, 35 percent smaller than the Agilent 8495/7 (26.5 GHz) family, allows for easy integration into instruments and ATE systems.

Superior RF Performance

  • Frequency range: dc to 26.5 GHz
  • Insertion loss at 0 dB: 0.8 dB + 0.04 dB/GHz
  • Maximum SWR: 1.3 to 12.4 GHz, 1.7 to 26.5 GHz

Repeatability and Life Span

  • Repeatability life: 0.03 dB max 5 million cycles per section, typical
  • 5 million cycles per section

Power Handling

  • Maximum RF input power: 1 W (average), 50 W (peak)

Features

  • Connector type: choice of 3.5 mm female, 3.5 mm male/female
  • Programming connector: 10-pin sockets
List of Options
004 3.5 mm female, 3.5 mm female
011 5 Vdc supply voltage (nominal)
015 15 Vdc supply voltage (nom)
024 24 volt solenoids supply volgate
104 3.5mm (m) connector, 3.5mm (f) connector
890 Calibration Data
C05 3.5 mm connector male to 3.5 mm connector male
H89 S21 measurement data on 3.5 diskette
R-50C-011-3 Agilent Calibration - 3 years
R-50C-011-5 Agilent Calibration - 5 years
R-50C-021-3 ANSI Z540-1-1994 Calibration - 3 years
R-50C-021-5 ANSI Z540-1-1994 Calibration - 5 years
UK6 Commercial cal. certificate w/ test data