Variable size stroke adjustment, allows the user to vary the volume and pressure per stroke
Integral pressure adjustment vernier knob for fine pressure adjustment
Directly connection to pressure modules consolidates two separate pieces into one when performing pressure calibration and testing
Two ¼ inch NPT pressure ports, one for reference (pressure module or gauge), one for UUT
Compatible with optional Fluke-700PRV pressure relief valve designed to protect pressure modules from overpressure damage
Note: The user must provide a hose with appropriate end fittings to apply pressure from the UUT port to the calibration unit under test such as the Fluke-700HTH-1 hose kit.
Includes:
¼ inch NPT to ¼ inch NPT adapter to connect the reference gauge port on the pump body to the 1/4 NPT female fitting on Fluke 700P29, P30, and P31 pressure modules