Stanford Research Systems SR865

2 MHz lock-in amplifier
Contact us for pricing and availability.
Description
  • 1 mHz to 2 MHz (operates to 2.5 MHz)
  • Low noise voltage and current inputs
  • 1 µs to 30 ks time constants
  • High bandwidth outputs
  • Touchscreen data display - large numeric results, chart recordings, & FFT displays
  • 10 MHz timebase input and output
  • GPIB, RS-232, Ethernet and USB
  • HDMI video output 
Signal Channel
Voltage inputsSingle-ended (A) or differential (A-B)
Sensitivity (output scale)1 nV to 1 V (voltage input)
1 fA to 1 µA (current input)
Voltage Input range10 mV to 1 V (peak).
Current input range1 µA or 10 nA (peak)
Max input before overload1 V (peak) or 1 µA (peak)
Input impedance
     Voltage input10 MΩ + 25 pF, AC (>1 Hz) or DC coupled
     Current input1 kΩ or 100 Ω to virtual ground
Gain accuracy±1 % (<200 kHz), ±2 % (to 2 MHz). Signal amplitude must be less than 30% of input range.
Noise2.5 nV/√Hz at 1 kHz (10 mV input range, typ.)
Harmonic distortion-80 dB (<100 kHz), -60 dB (>100 kHz)
CMRR90 dB at 1 kHz (DC coupled)
Dynamic reserve120 dB (typ)
Reference Channel
Frequency range0.001 Hz to 2 MHz, specified. (operates to 2.5 MHz)
Timebase10 MHz In/Out (phase locks the internal frequency to other SR865s.)
Input impedance1 MΩ or 50 Ω
Phase setting resolution(360/232) degrees
Phase noise
     Int. reference<0.0001° rms at 1 kHz (100 ms, 12 dB/oct)
     Ext. reference  (typ)<0.001° rms at 1 kHz (100 ms, 12 dB/oct)
Phase drift (typ)

 

Sine Out to signal In (200 mVrms) 
<0.002°/°C below 20 kHz (DC coupled)
<0.02°/°C below 200 kHz
<0.2°/°C below 2 MHz
Harmonic detectionDetect at N × fref where N<99 and (N × fref ) < 2 MHz
Dual F referenceDetect at fdual =¦fint fext¦
All frequencies must be less than 2 MHz
Chopper referenceSR865 drives SR540 Chopper (via Aux Out 4) to lock the chopper to fint
Demodulator
DC stabilityDigital output values have no offset drift
Time constants1 µs to 30 ks
Low pass filtersTypical RC-type filters or advanced Gaussian/Phase-Linear filters
Filter slope6, 12, 18 or 24 dB/oct rolloff
Synchronous filterAvailable below 4 kHz
Harmonic rejection-80 dB
Low latency outputRear-panel BlazeX output with <2 µs delay (plus LPF rise/fall times).
Internal Oscillator
Frequency1 mHz to 2 MHz, specified (operates to 2.5 MHz)
Frequency accuracy25 ppm + 30 µHz (with internal timebase)
External timebase10 MHz timebase input/output on rear panel
Frequency resolution6 digits or 0.1 mHz (whichever is greater)
Sine Output
OutputsDifferential or single-ended
Output impedance50 Ω source
Amplitude1 nVrms to 2 Vrms (specified amplitude is differential into 50 Ω loads)
Output amplitude is halved when used single-ended
Output amplitude is doubled into a high impedance load
Amplitude resolution3 digits or 1 nV, whichever is greater
DC offset±5 V, differential or common mode
Offset resolution3 digits or 0.1 mV, whichever is greater
Output limit±6 V, sum of DC offset and peak amplitude
Amplitude stability50 ppm/°C (typ)
SyncLogic level sync on rear panel (via BlazeX output)
Data
Data channels4 data channels are displayed and graphed (green, blue, yellow, orange)
Data sourcesEach data channel can be assigned any of these data sources:
X, Y, R, Θ, Aux In 1 to 4, Aux Out 1 to 2, X noise, Y noise, Sine Out Amplitude, 
Sine Out DC Level, Reference Phase, or Reference Frequency
Data historyAll data sources are continually stored at all chart display time scales.
The complete stored history of any data source can be displayed at any time.
OffsetX, Y and R may be offset up to ±999 % of the output scale
ExpandX, Y and R may be expanded by ×10 or ×100
RatioX and Y may be ratioed by Aux In 3. R may be ratioed by Aux In 4
Capture buffer1 Mpoints internal data storage. Store (X), (X and Y), (R and Θ) or (X, Y, R and Θ)
at sample rates up to 1.25 MHz. This is in addition to the data histories for the chart display.
Data streamingRealtime streaming of data, either (X), (X and Y), (R and Θ), or (X, Y, R and Θ) at sample rate to 1.25 MHz over Ethernet interface
ScanningOne of the following parameters may be scanned:
fint, Sine Out Amplitude, Sine Out DC Level, Aux Out 1 or 2.
FFT
SourceInput ADC, demodulator output, or filter output
Record lengthI1024 bins
AveragingExponential rms
Inputs and Outputs
CH1 outputProportional to X or R, ±10 V full scale thru 50 Ω
CH2 outputProportional to Y or Θ, ±10 V full scale thru 50 Ω
X and Y rear-panel outputsProportional to X and Y, ±10 V full scale thru 50 Ω
BlazeXLow latency output of X, ±2.0 V full scale or logic level reference sync output, either thru 50 Ω
Aux outputs4 BNC D/A outputs, ±10.5 V thru 50 Ω, 1 mV resolution
Aux inputs4 BNC A/D inputs, ±10.5 V, 1 mV resolution, 1 MΩ input
Trigger inputTTL input triggers storage into the internal capture buffer
Monitor outputAnalog output of the signal amplifier
HDMIVideo output to external monitor or TV, 640 × 480, 60 Hz
Timebase I/O1 Vrms, 10 MHz clock to synchronize internal reference frequency to other units
General
InterfacesIEEE488.2, RS-232, USB device and Ethernet.
USB flashFront-panel slot for USB flash storage of screen shots and data
Preamp power9-pin D connector to power SRS preamps
Power60 W, 100/120/220/240 VAC, 50/60 Hz
Dimensions17" × 5.25" × 17" (WHL)
Weight22 lbs.
WarrantyOne year parts and labor on defects in materials and workmanship

Request a Quote